13. ERROR DISPLAY
Error display Description
Above measuring range error
The viscosity value exceeds the upper limit of the viscosity
measuring range (SV-10: 11.99 Pas, SV-100: 119.9 Pas).
The viscosity of the sample can not be measured.
This error may occur when the display unit is not connected to
the main unit.
Below measuring range error
The viscosity value is below the lower limit of the viscosity
measuring range (SV-10: 0.30 mPas, SV-100: 0.90 Pas)
The viscosity of the sample can not be measured.
This error may occur when the display unit is not connected to
the main unit.
The power supply for the internal clock is depleted.
Press any key to enter the clock correction mode. Setting the
clock enables the viscometer to be used temporarily.
If the error occurs frequently, contact the local A&D dealer for
repair.
Internal IC error
Turn the power off. Then, turn the power on again. If this does
not release the error, contact the local A&D dealer for repair.