30
7-2 Details of the Function Items
Function item Parameter Description
Cladj
Date/Time
Sets the order of the date (YMD,MDY,DMY) and the
date/time.
Cond 0 Follows the viscosity changes quickly. (Prone to vibration)
Condition
1
•
2 Follows the viscosity changes slowly. (Stable values)
Unit 0 • mPa⋅s
Unit upon power-on 1 • Pa⋅s
2 cP
3 P
°C
4 mPa⋅s
5 Pa⋅s
6 cP
7
Viscos
-ity
P
Temper-
ature
°F
Factory setting:
SV-10=0
SV-100=1
pnt 0 • Dot
Decimal point 1 Comma
With "Comma" selected, the
separator for CSV format will
be ";" (semicolon).
fnc 0 • Switches viscosity units.
MODE key function
during measurement
1
Switches between the temperature display and the
measurement elapsed time display.
prt 0 • Key mode
Press the PRINT key to output data.
Data output mode 1 Auto print mode
Outputs automatically when
the STOP key ends the
measurement.
2 Stream mode
Continuous output during
measurement. Outputs the viscosity
only when D.P. format is selected.
type 0 A&D standard format For AD-8121B MODES 1&2
Data output format 1 • D.P. format For AD-8121B MODE 3
2 CSV format For a personal computer
3 RsVisco format For graphing program RsVisco
5-at 0 No output
Measurement
elapsed time output
1 • Output
Available only for D.P.
format
5-td 0 No output
Date/time output 1 • Output
Available only for D.P. and
CSV formats
5-ed 0 No output
Other output 1 Outputs remarks.
2 •
Outputs remarks, Device ID
information and signature.
Available only for D.P. format
3 Outputs ID number. Available only for CSV format
pU5e 0 No pause
Pause at data
output
1 • Pause (Approx. 2 seconds)
erfnc 0 • Usually use this parameter.
Reserved |
7
id
Device ID number
Set the device ID number.
With "5-ed", the device ID information
is added to the measurement data.
Clr
Initialization
Restores the function settings and calibration data to the
factory setting.
• Factory setting
Komentarze do niniejszej Instrukcji