40
7-4 Data Output Format Examples
7-4-1 A&D Standard Format
Used with the printer MODE 1 or MODE 2 when the optional compact printer, AD-8121B is
connected. Only the viscosity value is output.
SV-10 output format example
Viscosity
unit
Display Output format Remarks
LLLLL mPa・s OL,-99999999mPs Below measuring range error
0.30mPa・s ST,+00000.30mPs
10.0 mPa・s ST,+00010.00mPs The digit of 0.01mPa⋅s is always zero.
100 mPa・s ST,+00100.00mPs The digits of 0.01mPa⋅s and
0.1mPa⋅s are always zero.
1.00 Pa・s ST,+01000.00mPs For 1000mPa⋅s or greater, the
displayed unit is Pa⋅s, but the
output unit remains mPa⋅s.
The digits of 0.01mPa⋅s, 0.1mPa⋅s
and 1mPa⋅s are always zero.
mPa・s
H Pa・s OL,+99999999mPs Above measuring range error
L Pa・s OL,-99999999Pas Below measuring range error
0.0003 Pa・s ST,+000.0003Pas
0.0100 Pa・s ST,+000.0100Pas
0.100 Pa・s ST,+000.1000Pas The digit of 0.0001Pa⋅s is always
zero.
1.00 Pa・s ST,+001.0000Pas The digits of 0.0001Pa⋅s and
0.001Pa⋅s are always zero.
Pa・s
H Pa・s OL,+99999999Pas Above measuring range error
L cP OL,-99999999 cP Below measuring range error
0.30 cP ST,+00000.30 cP
10.0 cP ST,+00010.00 cP
The digit of 0.01cP is always zero.
100 cP ST,+00100.00 cP
The digits of 0.01cP and 0.1cP are
always zero.
10.0 P ST,+01000.00 cP For 1000 cP or greater, the displayed
unit is P, but the output unit remains cP.
The digits of 0.01cP, 0.1cP and
1cP are always zero.
cP
H P OL,+99999999
cP Above measuring range error
L P OL,-99999999 P Below measuring range error
0.0030 P ST,+000.0030 P
0.100 P ST,+000.1000 P
The digit of 0.0001P is always
zero.
1.00 P ST,+001.0000 P
The digits of 0.0001P and 0.001P
are always zero.
10.0 P ST,+010.0000 P
The digits of 0.0001P, 0.001P and
0.01P are always zero.
P
H P OL,+99999999
P Above measuring range error
: Space (ASC 20h)