Ad-weighing MPA-1200 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Ad-weighing MPA-1200. A&D Weighing MPA-1200 User Manual [en] Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - 1WMPD4002874B

MPA-10/MPA-20/MPA-200/MPA-1200 Single Channel Electronic Pipette INSTRUCTION MANUAL 1WMPD4002874B

Strona 2

86. PACKING CONTENTS AND NAME OF ITEMS Confirm that the following contents are all included. ○ Electronic pipette MPA-10 / 20 / 200 / 1200 (Any o

Strona 3 - CONTENTS

9 Should the pipette arrive damaged or an accessory be missing, contact the nearest A&D dealer. Note The accessories included wi

Strona 4

107. PREPARATION BEFORE USE 7-1 Installing the battery 1. Remove the battery cover (2) by sliding it upward while pressing and holding the batte

Strona 5 - 1. FOR SAFE USE

117-2 Recharging the battery When purchasing the instrument, the battery does not have a full charge. For initial use, first charge the battery full

Strona 6 - 3. FEATURE

127-3 Exchange selectable power plug The A-Type power supply plug is originally attached to the AC adapter. Please change the power supply plug to

Strona 7 - 4. COMPLIANCE

13Operation Turning the power off manually (The display is example.) Press and hold the Operation key on the pipette for at least five seconds unt

Strona 8

147-5 Parts names The following shows the each name of electronic pipette When confirming LCD, refer to “8-1 Display and functions” for details.

Strona 9 - 5. MPA FUNCTION

158. NAME AND FUNCTIONS OF DISPLAY AND KEYS 8-1 Display and functions Symbols Descriptions System setting mode Used to set up functions

Strona 10

16 Symbols Descriptions Aspirating and dispensing speed display Shows the speed level when aspirating or dispensing the liquid.▲ blinks

Strona 11

179. FUNCTION AND HOW TO USE The MPA series have three modes, the standard mode (AUTO), multiple dispensing mode (MD) and the mixing mode (MIX). 9

Strona 12 - 7-1 Installing the battery

WARNING DEFINITIONS The warnings described in this manual have the following meanings: DANGER An imminently hazardous situation which, if not avoi

Strona 13 - 7-2 Recharging the battery

183) Operating the standard mode [ 1 ] Put the tip end into the liquid to be dispensed and press the Operation key on the pipette to aspirat

Strona 14

199-2 Multiple dispensing mode (MD) 1) Operation of the multiple dispensing mode This is a suitable function to dispense the same liquid continuousl

Strona 15

20[ 7 ] Press the Enter key to confirm the dispensing count. If you would like to change dispensing amount or dispensing count, press the En

Strona 16 - 7-5 Parts names

21[ 4 ] Repeat the operation described above with the dispensing count set. When the set count of dispensing operations is completed a buzzer will so

Strona 17 - 8-1 Display and functions

22[ 3 ] Press the Enter key to select mixing mode (MIX). [ 4 ] Press the Up or Down key to set mixing count (One set with aspirating an

Strona 18

23[ 3 ] The pipette executes the set count of aspirating-dispensing cycles by approx. 2/3 of the set mixing volume. [ 4 ] Press the

Strona 19 - 9. FUNCTION AND HOW TO USE

24 [ 2 ] Press the Up or Down key to select “SYS”. [ 3 ] Press the Enter key to enter the system setting mode. [ 4 ] By pressing the

Strona 20

259-5 Program setting mode The nine programs can be saved in the program memory built into the pipette (PROG 01 to 09). By saving a frequently used

Strona 21

261 ) Setting the reverse mode (The display is example.) [ 1 ] Press the Back key

Strona 22

272) Operating the reverse mode [ 1 ] Press the Operation key on the pipette to aspirate the liquid. [ 2 ] Press the O

Strona 23 - 9-3 Mixing mode(MIX)

1 CONTENTS 1. FOR SAFE USE... 3 1-1 Precauti

Strona 24

289-8 Total discharge function Pressing and holding the Operation key expels all of the liquid remaining in the tip. This function is useful wh

Strona 25

2911. CALIBRATING THE PIPETTE USING AN ELECTRONIC BALANCE The MPA series provide user with a dispensing volume calibration function. Using this fu

Strona 26

30[ 3 ] Press the Enter key to enter μL calibration mode. [ 4 ] Press the Up or Down key to alter the value to the actual dispensed

Strona 27 - 9-5 Program setting mode

31[ 3 ] Press the Up or Down key to select volume calibration data resetting mode. [ 4 ] Press the Enter key to enter volume calibrati

Strona 28

32With theμL unit selected, the Noticemark is turned off while the μLunit mark lights up. With the mg unit selected, the Notice markblinks while th

Strona 29 - 9-7 Blowout function

33Method for mg calibration 1. Select the mg unit beforehand. (Refer to “11-3 Dispensing in a unit of weight (in mg unit)”) 2. Set the dispensing a

Strona 30

3412. ADJUSTING HEIGHT OF THE TIP EJECTOR A height of the tip ejector can be adjusted so that it can match the conditions of how the tip used was

Strona 31

3513-1 Autoclave By removing the lower part of pipette from the pipette, it can be placed in an autoclave. Condition for autoclave: Lower part of pi

Strona 32

3614. TROUBLE SHOOTING The electronic pipette has a self-check function. When there is an error in these results, the pipette displays a following

Strona 33

3715. WHEN REQUESTNG REPAIRE The pipette requires repair if an error occurs and cannot be corrected by following the troubleshooting methods provid

Strona 34

217. LIST OF ITEMS SOLD SEPARATELY(DISPOSABLE ITEMS)...39 17-1 Stands and hanger...

Strona 35

3816. SPECIFICATION MPA-10 MPA-20 MPA-200 MPA-1200 Volume range 0.5 to 10.0μL 2.0 to 20.0μL 10 to 200μL 100 to 1200μL Volume 1.0μL 10.0μL 2.0μL 20

Strona 36

3917. LIST OF ITEMS SOLD SEPARATELY(DISPOSABLE ITEMS) 17-1 Stands and hanger Applicable pipette Number Name MPA-10 MPA-20 MPA-200 MPA-1200AX-ST-

Strona 37 - 13-1 Autoclave

4017-2 Tips and accessories Applicable pipette Number Name MPA-10 MPA-20 MPA-200 MPA-1200AX-BOX-200A Tip box with locking parts *4 *5 (10/20/200μL)

Strona 38 - 14. TROUBLE SHOOTING

41○AX-BOX-1200B Tip box ○AX-CART-1200 Tip cartridge (1200μL) Consultation) The tip is marked with lines to act as guidelines for the a

Strona 39 - 15. WHEN REQUESTNG REPAIRE

4217-3 Disposable items (User replaceable) Applicable pipette Number Name MPA-10 MPA-20 MPA-200 MPA-1200AX-LOW-10 Lower part (10μL) ○ AX-LOW-

Strona 40 - 16. SPECIFICATION

4317-4 Others - Leak tester AD-1690 Leakage within the pipette can be easily checked. - Pipette accuracy tester BM-20/22(BM-014 attached)

Strona 42 - 17-2 Tips and accessories

45Attestation of contamination removal Please fill in the following items when you send a pipette for repair. Model name: Seri

Strona 43

3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, IN

Strona 44

31. FOR SAFE USE 1-1 Precautions on using the pipette DANGER - This instrument is not an explosion proof instrument. Do not use the pipette

Strona 45 - 17-4 Others

4 CAUTION - Do not use a leaking battery. - Because the battery body may become hot when using the pipette continuously for a long time, take care no

Strona 46

54. COMPLIANCE Compliance with FCC Rules Please note that this device generates, uses and can radiate radio frequency energy. This device has been

Strona 48

75. MPA FUNCTION □ The pipette has three modes where advantages of electromotion are utilized. (Refer to “9. FUNCTION AND HOW TO USE”) - Standard

Powiązane modele: MPA-200 | MPA-20 | MPA-10 |

Komentarze do niniejszej Instrukcji

Brak uwag