ふ MS-70 MX-50 MF-50 ML-50 WM+PD4000477D

Strona 2

MS-70 / MX-50 / MF-50 / ML-50 8 4.4.4.4. Outline and Outline and Outline and Outline and FeaturesFeaturesFeaturesFeatures The moisture an

Strona 3 - ContentsContents

MS-70 / MX-50 / MF-50 / ML-50 9 The analyzer is equipped with a serial interface as standard. It can be connected to a printer or computer.

Strona 4

MS-70 / MX-50 / MF-50 / ML-50 10 5.5.5.5. Packing List and Names of each partPacking List and Names of each partPacking List and Names of each pa

Strona 5

MS-70 / MX-50 / MF-50 / ML-50 11 All AccessoriesAll AccessoriesAll AccessoriesAll Accessories Caution Please confirm tha

Strona 6

MS-70 / MX-50 / MF-50 / ML-50 12 5.1.5.1.5.1.5.1. Display and KeysDisplay and KeysDisplay and KeysDisplay and Keys Name State and Me

Strona 7

MS-70 / MX-50 / MF-50 / ML-50 13 #1 ML-50 can select "standard drying " and "quick drying" only. #2 ML-50 does not display &quo

Strona 8

MS-70 / MX-50 / MF-50 / ML-50 14 6.6.6.6. PreparationPreparationPreparationPreparationssss 6.1.6.1.6.1.6.1. Installing the AnalyzerInsta

Strona 9

MS-70 / MX-50 / MF-50 / ML-50 15 6.2.6.2.6.2.6.2. Setting the Clock and CSetting the Clock and CSetting the Clock and CSetting the Clock and Calend

Strona 10

MS-70 / MX-50 / MF-50 / ML-50 16 6.3.6.3.6.3.6.3. Proper Operation for Precision MeasurementProper Operation for Precision MeasurementProper Oper

Strona 11

MS-70 / MX-50 / MF-50 / ML-50 17 6.3.3.6.3.3.6.3.3.6.3.3. The Glass Fiber SheetThe Glass Fiber SheetThe Glass Fiber SheetThe Glass Fiber Sheetssss

Strona 12

1.1.1.1. This manual and MarksThis manual and MarksThis manual and MarksThis manual and Marks All safety messages are identified by the fol

Strona 13

MS-70 / MX-50 / MF-50 / ML-50 18 7.7.7.7. Measurement ProcedureMeasurement ProcedureMeasurement ProcedureMeasurement Procedure 7.1.7.1.7.1.7.

Strona 14 - 5td, quc, U-a, U-t, U-m

MS-70 / MX-50 / MF-50 / ML-50 19 1. Turn on the analyzer. The gram unit (of weighing mode) is displayed. Enter the Standard ModeEnter the Stan

Strona 15

MS-70 / MX-50 / MF-50 / ML-50 20 11. Open the heater cover. Put in a sample using the level indicator. Caution The sample needs to be m

Strona 16

MS-70 / MX-50 / MF-50 / ML-50 21 7.2.7.2.7.2.7.2. Quick Mode OperationQuick Mode OperationQuick Mode OperationQuick Mode Operation The quic

Strona 17 - Cl adj

MS-70 / MX-50 / MF-50 / ML-50 22 1. Turn on the analyzer. The gram unit (of weighing mode) is displayed. Enter the Quick ModeEnter the Quick

Strona 18 - 0.000g

MS-70 / MX-50 / MF-50 / ML-50 23 11. Open the heater cover. Put in a sample using the level indicator. Caution The sample needs to be mor

Strona 19

MS-70 / MX-50 / MF-50 / ML-50 24 7.3.7.3.7.3.7.3. Program NumberProgram NumberProgram NumberProgram Number The measurement conditions of

Strona 20

MS-70 / MX-50 / MF-50 / ML-50 25 8.8.8.8. MeasurementMeasurementMeasurementMeasurement Programs Programs Programs Programs 8.1.8.1.8.1.8.1. Li

Strona 21

MS-70 / MX-50 / MF-50 / ML-50 26 8.1.1.8.1.1.8.1.1.8.1.1. ACCURACY of the Standard Mode and Quick ModeACCURACY of the Standard Mode and Quick Mode

Strona 22

MS-70 / MX-50 / MF-50 / ML-50 27 8.1.2.8.1.2.8.1.2.8.1.2. Analyzing mode of the Automatic ModeAnalyzing mode of the Automatic ModeAnalyzing mode of

Strona 23

MS-70 / MX-50 / MF-50 / ML-50 1 ContentsContentsContentsContents 1. Safety and Compliance...

Strona 24

MS-70 / MX-50 / MF-50 / ML-50 28 How to Select quick mode (Extracts)How to Select quick mode (Extracts)How to Select quick mode (Extracts)How to Se

Strona 25

MS-70 / MX-50 / MF-50 / ML-50 29 8.2.8.2.8.2.8.2. Procedures to Store a Measurement ProgramProcedures to Store a Measurement ProgramProcedures to S

Strona 26

MS-70 / MX-50 / MF-50 / ML-50 30 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SEL

Strona 27 - STOP key. Max

MS-70 / MX-50 / MF-50 / ML-50 31 Set Set Set Set the Minimum Scale Value of the Gram Displaythe Minimum Scale Value of the Gram Displaythe Minimum S

Strona 28 - HI , MID. or LO

MS-70 / MX-50 / MF-50 / ML-50 32 8.2.2.8.2.2.8.2.2.8.2.2. Ramp DryingRamp DryingRamp DryingRamp Drying Standard Standard Standard Standard Mode

Strona 29

MS-70 / MX-50 / MF-50 / ML-50 33 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELEC

Strona 30 - quc using the ↓ or ↑ key

MS-70 / MX-50 / MF-50 / ML-50 34 Store Parameters and Finish the OperationStore Parameters and Finish the OperationStore Parameters and Finish the

Strona 31

MS-70 / MX-50 / MF-50 / ML-50 35 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELEC

Strona 32 - U-a

MS-70 / MX-50 / MF-50 / ML-50 36 8.2.3.8.2.3.8.2.3.8.2.3. Step DryingStep DryingStep DryingStep Drying Standard Mode, Automatic Mode or Standar

Strona 33

MS-70 / MX-50 / MF-50 / ML-50 37 Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELEC

Strona 34

MS-70 / MX-50 / MF-50 / ML-50 2 8. Check Function...41 8.1. Self-Check Fu

Strona 35

MS-70 / MX-50 / MF-50 / ML-50 38 Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Val

Strona 36

MS-70 / MX-50 / MF-50 / ML-50 39 1. Display gram unit (of the weighing mode). Select a Program Number to Edit the Measurement Select a Progra

Strona 37

MS-70 / MX-50 / MF-50 / ML-50 40 Set tSet tSet tSet the he he he Measurement Measurement Measurement Measurement UnitUnitUnitUnit 11. Select

Strona 38

MS-70 / MX-50 / MF-50 / ML-50 41 9.9.9.9. Check FunctionCheck FunctionCheck FunctionCheck Function 9.1.9.1.9.1.9.1. SelfSelfSelfSelf----Check

Strona 39

MS-70 / MX-50 / MF-50 / ML-50 42 9.2.9.2.9.2.9.2. Test Sample (STest Sample (STest Sample (STest Sample (Sodium odium odium odium TTTTartrate art

Strona 40 - "0."0"1%

MS-70 / MX-50 / MF-50 / ML-50 43 10.10.10.10. Connecting to a PrinterConnecting to a PrinterConnecting to a PrinterConnecting to a Printer

Strona 41 - U-t of the timer mode with

MS-70 / MX-50 / MF-50 / ML-50 44 10.1.10.1.10.1.10.1. Print samplesPrint samplesPrint samplesPrint samples 10.1.1.10.1.1.10.1.1.10.1.1. Exampl

Strona 42 - "0"0"1%

MS-70 / MX-50 / MF-50 / ML-50 45 10.1.2.10.1.2.10.1.2.10.1.2. Example To Print Selected ItemsExample To Print Selected ItemsExample To Print Selecte

Strona 43 - CH pa55

MS-70 / MX-50 / MF-50 / ML-50 46 Print ExamplePrint ExamplePrint ExamplePrint Example üüüüüüüüüüüAü&üD Factory MODELüüüüüüMX-50 Product S/

Strona 44 - 15.80%

MS-70 / MX-50 / MF-50 / ML-50 47 10.1.3.10.1.3.10.1.3.10.1.3. ExplExplExplExplanation for Print Itemanation for Print Itemanation for Print Itemanat

Strona 45

MS-70 / MX-50 / MF-50 / ML-50 3 2.2.2.2. Safety Safety Safety Safety and Complianceand Complianceand Complianceand Compliance WARNING Do no

Strona 46 - Signature. Refer to 10.1.3

MS-70 / MX-50 / MF-50 / ML-50 48 """"Measurement Measurement Measurement Measurement Unit"Unit"Unit"Unit"

Strona 47

MS-70 / MX-50 / MF-50 / ML-50 49 11.11.11.11. Connecting to a ComputerConnecting to a ComputerConnecting to a ComputerConnecting to a Computer

Strona 48

MS-70 / MX-50 / MF-50 / ML-50 50 11.1.11.1.11.1.11.1. RSRSRSRS----232C Serial Interface232C Serial Interface232C Serial Interface232C Serial Inter

Strona 49

MS-70 / MX-50 / MF-50 / ML-50 51 11.2.11.2.11.2.11.2. Output FormatOutput FormatOutput FormatOutput Format In Case of Format omitted Temperature

Strona 50

MS-70 / MX-50 / MF-50 / ML-50 52 In Case of Format included Temperature Data (Function Table In Case of Format included Temperature Data (Function

Strona 51

MS-70 / MX-50 / MF-50 / ML-50 53 12.12.12.12. Data Memory FunctionData Memory FunctionData Memory FunctionData Memory Function The data mem

Strona 52

MS-70 / MX-50 / MF-50 / ML-50 54 12.1.2.12.1.2.12.1.2.12.1.2. Output All Data at One TimeOutput All Data at One TimeOutput All Data at One TimeOut

Strona 53

MS-70 / MX-50 / MF-50 / ML-50 55 13.13.13.13. CalibrationCalibrationCalibrationCalibration The moisture content is calculated with a ratio

Strona 54

MS-70 / MX-50 / MF-50 / ML-50 56 13.2.13.2.13.2.13.2. CalibCalibCalibCalibration of the Weighting Sensorration of the Weighting Sensorration of th

Strona 55 - data "1

MS-70 / MX-50 / MF-50 / ML-50 57 Calibration RCalibration RCalibration RCalibration Reporteporteporteport Example for the Weighi Example for the Weig

Strona 56

MS-70 / MX-50 / MF-50 / ML-50 4 Compliance with FCC RulesCompliance with FCC RulesCompliance with FCC RulesCompliance with FCC Rules Please

Strona 57 - Cl adj

MS-70 / MX-50 / MF-50 / ML-50 58 13.3.13.3.13.3.13.3. Calibration of Drying TemperatureCalibration of Drying TemperatureCalibration of Drying Temp

Strona 58

MS-70 / MX-50 / MF-50 / ML-50 59 10. After fifteen minutes, the buzzer sounds and the display blinks "1"6"0°C . Adjust the bl

Strona 59 - } Calibration type

MS-70 / MX-50 / MF-50 / ML-50 60 14.14.14.14. Function TableFunction TableFunction TableFunction Table The function table can store the f

Strona 60 - 100℃ "9"7℃

MS-70 / MX-50 / MF-50 / ML-50 61 14.1.1.14.1.1.14.1.1.14.1.1. OperationOperationOperationOperation 1. Display the gram unit (of the weighing m

Strona 61

MS-70 / MX-50 / MF-50 / ML-50 62 15.15.15.15. MaintenanceMaintenanceMaintenanceMaintenance Turn off the power switch and remove power cor

Strona 62 - ENTER key

MS-70 / MX-50 / MF-50 / ML-50 63 15.2.15.2.15.2.15.2. Replacement of the Halogen LampReplacement of the Halogen LampReplacement of the Halogen LampR

Strona 63 - dp "1

MS-70 / MX-50 / MF-50 / ML-50 64 15.3.15.3.15.3.15.3. Factory SettingsFactory SettingsFactory SettingsFactory Settings This function can s

Strona 64

MS-70 / MX-50 / MF-50 / ML-50 65 15.4.15.4.15.4.15.4. TroubleshootingTroubleshootingTroubleshootingTroubleshooting 1111....In the Case thatIn th

Strona 65

MS-70 / MX-50 / MF-50 / ML-50 66 15.5.15.5.15.5.15.5. Error MessageError MessageError MessageError Message CH no Internal Error An inter

Strona 66

MS-70 / MX-50 / MF-50 / ML-50 67 16.16.16.16. SpecificationsSpecificationsSpecificationsSpecifications MS-70 MX-50 MF-50 ML-50 Measurement met

Strona 67

MS-70 / MX-50 / MF-50 / ML-50 5 3.3.3.3. PrecPrecPrecPrecautionsautionsautionsautions 3.1.3.1.3.1.3.1. Installing the AnalyzerInstalling the A

Strona 68

MS-70 / MX-50 / MF-50 / ML-50 68 MS-70 MX-50 MF-50 ML-50 Operation environment 5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation) Sample

Strona 69

MS-70 / MX-50 / MF-50 / ML-50 69 16.2.16.2.16.2.16.2. Accessories and Accessories and Accessories and Accessories and PeripheralPeripheralPeripheral

Strona 70

MS-70 / MX-50 / MF-50 / ML-50 70 17.17.17.17. IndexIndexIndexIndex %/min ---------------------------- 18, 21, 26, 28 , ENTER key -----------

Strona 71

MS-70 / MX-50 / MF-50 / ML-50 71 ISO ------------------------------- 43, 56, 59, 60 Key mode-------------------------------------- 60 Keys-----------

Strona 72

MS-70 / MX-50 / MF-50 / ML-50 72 MEMO

Strona 73

MS-70 / MX-50 / MF-50 / ML-50 73

Strona 74

MS-70 / MX-50 / MF-50 / ML-50 6 3.2.3.2.3.2.3.2. During UseDuring UseDuring UseDuring Use ------------Caution for MCaution for MCaution for M

Strona 75

MS-70 / MX-50 / MF-50 / ML-50 7 Caution for Heating (Drying). When the drying temperature is set to 200°C and measurement is started, the the